10. RAD-IO2 CALIBRATION

  • Enable Calibration:

Goto Settings Menu setting

_images/setting_status.jpg

Change calibration option to on

_images/calibration_optn.jpg

Calibration page should now be available

_images/calibration.jpg

Click Read to get current calibration constants

_images/calibration_read.jpg

Equipment needed for calibration:

TC and AIN:

Fluke 7562A or equivalent

AOUT

6 ½ digit multimeter such as Keysight 34465A or Keithley DMM6500 or equivalent

_images/calibration_dash.jpg

Read: Reads current calibration data

Store: Writes displayed calibration data to device

Export: Saves copy of current calibration data

Import: Loads saved calibration data

Clear: Clears existing calibration data, Will reset so all calibrations points have zero error.

Manual: Requires the user to enter the calibration value

Interactive: Reads actively reads input. Pressing enter or tab will advance to the next bank then next point. This is the fastest way to calibrate a device.